METHOD

Features of interlaboratory comparison methods when measuring vibroacoustic parameters

Sterlikov AV1, Kurilenko YuV2, Voronkov AA2
About authors

1 Research and Technical Center of Radiation-Chemical Safety and Hygiene of Federal Medical-Biological Agency, Moscow, Russia

2 Tsifrovye Pribory Ltd, Moscow, Russia

Correspondence should be addressed: Alexander V. Sterlikov
Shchukinskaya, 40, Moscow, 123182; ur.liam@vokilretsa

About paper

Author contribution: all authors contributed to data acquisition, data compilation, and manuscript writing.

Received: 2021-05-27 Accepted: 2021-06-11 Published online: 2021-06-18
|

The results of external quality control in the form of interlaboratory comparisons (ILCs) are an important criterion of the testing laboratory competence. The need for participation in ILCs is one of the requirements for testing laboratories codified in the law [1, 2]. Control over compliance with these requirements is exercised by both Rosaccreditation, and executive authorities in charge of accredited organizations [3, 4].

Among tests, studies, and measurements performed in order to evaluate compliance to hygiene standards, there are many direct measurements of physical factors, such as acoustic noise and vibration. Over recent years there has been an upward trend in the number of those [5].

Conducting ILCs when measuring physical factors, affecting humans in their living environment, such as acoustic noise and vibration, has a number of features. The main feature is that the object, the properties of which are being measured in the laboratory, cannot be sent to participants for measurement. Working place and residential development area can be considered the examples of such objects. Thus, the model objects are used for proficiency testing (OPT). It therefore seems necessary to perform ILC at the premises of provider ensuring availability of OPT and stability of OPT parameters. The other feature of noise and vibration measurement is impossibility of simultaneous measurement of one object by several laboratories, i.e. impossibility of conducting ILCs with the use of parallel programs. That is why the ILC providers use sequential programs with sequential access to OPT being the source of the measured factor. Meanwhile, the measured OPT parameters stability is controlled by reference laboratory [6] authorized by ILC provider.

Such practice of ILCs has been carried out since 2016 within the framework of the PHYSFACTOR-TEST voluntary certification system, as well as by other ILC providers [7, 8]. The method for carrying out and arranging ILCs when performing direct measurements has been reported [9].

The presence of OPT with appropriate characteristics is a key requirement to performing ILC using the approach described above.

Both natural and model objects may be used as OPTs. The model objects’ assessed parameters stability is ensured by specialized test benches (TBs).

When performing ILCs involving measurement of acoustic noise and vibration, it is difficult to use natural objects due to low stability, unpredictability of parameter changes during measurement, and impossibility of setting specific parameters.

The study was aimed to summarize the approaches to creation and the practice of using OPT based on physical simulation of acoustic noise sources, airborne ultrasound, and vibration for ILC, as well as to assess the effectiveness of using the TB developed as an OPT for ILC with measurement of acoustic noise and vibration.

METHODS

Analysis of the effectiveness of OPTs based on physical simulation of factors, the test benches (TB), was carried out based on the results of TB testing and standardization, as well as the results of ILCs performed with the use of those.

The basic requirements for TB are as follows:

  • the level of the factor generated should be in the range from the value three times (or by 10 dB) exceeding the detection threshold of the measurement systems, most commonly used by laboratories, to the level around the threshold limit value;
  • parameters of the studied feature (hereinafter the factor) should remain stable for at least 20–60 min;
  • susceptibility of TB performance to environmental factors should be kept to a minimum; there should also be a possibility of considering the impact of these factors;
  • it is necessary to ensure carrying out ILCs by all participants by way of a sequential access to TB;
  • it is necessary to take into account the measurement technique (direct measurement method) and the features of using the measurement system;
  • TB should adequately simulate the impact of the measured factor on the environment (working place, working area);
  • when constructing TB, it is necessary to strike a balance between thorough simulation of the factor impact on environmental objects and ensuring the stability and reproducibility of measurements;
  • TB should enable setting various modes and factor intensities;
  • TB should ensure reproducibility of levels being monitored when performing repeated measurements during the TB operation in the course of ILC;
  • when using TB, it is necessary to ensure resistance to externally acting interfering influences (interference immunity).

The results of using TB as an OPT were treated based on the following factors:

  • acoustic noise;
  • airborne ultrasound;
  • local vibration;
  • general machinery and household appliances vibration.

ILCs and measurements in the course of TB testing were carried out in accordance with the previously developed programs using the methods that had successfully passed metrological standardization. The ILC participants knew in advance the documents and were provided additional training prior to ILC.

All measurements were performed using the following measurement devices:

  • Ecophysika-110A system (OCTAVA-ElectronDesign; Russia);
  • АК-1000 acoustic calibrator (OCTAVA-ElectronDesign; Russia);
  • КВ-160 portable vibration reproducing device (vibrocalibrator) (OCTAVA-ElectronDesign. Russia).

RESULTS

Acoustical noise was the first type of ILC of physical factors. Currently, distribution of acoustic signal record as a reference for proficiency testing is being widely practiced. However, first, under this approach it could not be said that OPT is the same object the laboratory investigates. According to their areas of accreditation, laboratories perform measurements on various environmental objects (working areas, human living environment) or equipment testing. Under the approach discussed above, OPT is a speaker or digital recording; second, such approach in its current form has outlived its purpose and does not enable us to judge the ILC participant proficiency due to impossibility to perform quantitative comparison of the measurement results with the reproduced value. Furthermore, the content of measurement tasks does not comply with the changed regulatory framework.

In order to ensure interlaboratory comparison, we constructed TBs, playing back the noise with high stability. TBs consisted of waveform generator, power amplifier, and highquality full-range speaker. In order to assess the short-term stability, we used the reproduced by TB noise parameters of the model OPT, averaged during direct measurement. During the short periods, corresponding to typical ILC duration (up to 120 min), this parameter varied within the limit of ± 0.5 dB, which provided the short-term TB stability appropriate for ILC. Within the interval of several months with the same settings, the noise level reproduced by TB varied within the limit of ± 2 dB, which provided the long-term stability appropriate for continuous operation of TB.

The area of the facility, in which the measurement is performed, should be at least 40 m2, and the minimum linear size of the facility should be at least 2.5 m2. Standard room environment conditions for temperature, humidity, pressure, and air velocity should be maintained.

Test points should be selected 1 m away from the sound source at a height of 1.5 m from the surface of the floor. The reference point should be no closer than 1 m from the walls, window and door openings. Depending on the number of measured noise parameters and measurement program, the number of reference points may vary between 1 and 4.

Sensitivity of sound level meter (measurement system) was tested prior to measurement. The testing was performed in accordance with the sound level meter user guide. Sensitivity test was carried out using class 1 acoustic calibrator by GOST IEC 60942 having the verification certificate and approved by the sound level meter manufacturer.

During the measurements, the acoustic level (acoustic pressure) of background noise should be different from the acoustic level (acoustic pressure), provided by operating source, by not less than 10 dB. There should be no random noise in the facility.

The test duration was at least 1 min, and the number of measurements for each reference point was at least 3.

Measurement accuracy in accordance with the method used [10, 11] is characterized by expanded uncertainty of measurements with a coverage factor of 2, corresponding to 95% confidence interval. Expanded uncertainty of measured acoustic level with the method applied does not exceed 1.2 dB, and the expanded uncertainty of measured acoustic pressure does not exceed 2.2 dB.

More than 150 interlaboratory comparisons have been carried out.

Performing ILC with the acoustic level (acoustic pressure) measurement using the discussed TB met the need for correct comparison of the results obtained by the ILC participants and the reference laboratory, as well as for adequate  simulation of the real world conditions of acoustic noise measurement at the working place or other object. The proposed ILC method made it possible to set different levels for the factor to be measured. When conducting measurement, it is necessary to maintain the conditions required for prevention of erroneous results.

ILC with the airborne ultrasound level measurement was performed the same way. Design of TB, being the object for proficiency testing, was almost the same, however, the frequency of the acoustic signal played back was within the range of 11–22 kHz. High-frequency transducer with high pass filter having a cut-off frequency of 7–9 kHz was used to play back ultrasound signals. 

TB simulated the source of airborne ultrasound at the working place. The 1/3-octave-band acoustic pressure level was measured. TB made it possible to achieve the acoustic pressure levels of up to 85–87 dB.

The requirements for conducting measurements are generally the same as for acoustic noise parameters measurement. However, the important factor affecting the measurement results is pointing the sound level meter microphone at the source of ultrasound signal. In some microphones, it is necessary to consider the parameter adjustment for ultrasonic frequency range.

All acoustic measurements during ILC were performed with the microphone fixed in the reference point with a rack.

Interlaboratory comparisons, when vibration is measured by testing laboratories, are no less urgent and much more complex in terms of arranging and conducting. The measurement results are influenced by numerous factors: vibration source stability resulting from its design and operating mode, type and quality of the measurement system transducer fastening, mechanical interaction of the “worker–instrument–processed object” system, etc. That is why vibration measurement during simple simulation of the working process with the use of manual mechanized tool by the user would result in sharp discontinuities in the results, i.e. actually, there no short-term and long-term stability of measurements would be observed. This would make it difficult to compare these during interlaboratory comparisons. In addition, the user would be adversely affected by vibration. Even greater difficulties can be expected while ensuring measurement of general vibration, provided mostly by massive equipment and transport.

The circumstances described above have made it necessary to develop TBs that could be used as a part of OPT to measure the general vibration levels during ILCs.

Initially, the possibility of measuring local vibration with an instrument in idle or eccentric loading mode was studied.

However, in this case, stability proved insufficient. Thus, TB with automated specified mechanized tool vibration level maintenance using the proportional integral derivative (PID) controller was designed (fig. 1). Multifunction tool or drilling machine supported by hanger (bracing) was used as a manual mechanized tool.

The results of comparative vibration level measurements in the presence and absence of feedback are presented in fig. 2.

The data presented demonstrate that the system with feedback shows much more stable levels of local vibration generated for vibration acceleration. This justifies the increased complexity of TB for ILCs with measurement of local vibration.

To date, over 100 interlaboratory comparisons with the use of discussed TB have been performed.

When performing ILC with measurement of local vibration, a three-component vibration sensor was fixed on the instrument handle with a cubicle adapter in accordance with GOST 31192.2 [12]. Sensitivity axes of the vibration sensors were oriented in the Z, X and Y directions in accordance with the basicentric coordinate system of the human hand doing the work [13]. The absence of interference caused by electromagnetic fields and sensor cable movement were checked prior to measurement. Microclimatic conditions were consistent with the measurement system operation conditions.

The measurement was performed during a number of observation periods. At least four measurements of adjusted acceleration for each direction in each reference point were performed during each observation period. The duration of each test was at least 15 s.

Measurement accuracy in accordance with the method used [14] is characterized by expanded uncertainty of measurements with a coverage factor of 2, corresponding to 95% confidence interval. Expanded uncertainty of measurements performed using the method does not exceed 3.8 dB.

TB made it possible to achieve the level of vibration acceleration up to 135 dB.

The reported approach to performing ILC with the use of TB to generate stable vibration met the need for correct comparison of the results obtained by the ILC participants and the reference laboratory, as well as for adequate simulation of the real world conditions of local vibration measurement at the working place. The designed TB made it possible to set various levels of the factor to be measured, and ensured stable vibration generation.

A number of testing laboratories need to perform interlaboratory comparisons when measuring general machinery or household appliances vibration. Hygienic characteristics and standards of general machinery vibration, as well as the procedure of general machinery vibration measurement, differ from those of local vibration. General machinery vibration and household appliances vibration are the low-frequency vibration types, and are measured with the use of adapter, for example the floor standing platform mounted on the surface, on which the measurement is performed.

It is extremely difficult to perform ILCs with the measurement of general vibration on the basis of the physical object, therefore, it was necessary to develop the TB simulating the general vibration source. Such TB was developed. It was a platform sized 300 × 300 mm with a 3000 rpm electric motor fastened from below. The electric motor was operated in the eccentric loading mode, which resulted in vibration of the bench. In order to ensure stability and set the level of vibration acceleration, electric motor was included in the automatic control loop with vibration acceleration feedback, as described above. When performing measurement, the measurement system accelerometer was mounted on the platform surface. The platform itself was placed on a shock absorbing material.

TB made it possible to achieve vibration acceleration levels of up to 130 dB. During TB standardization, it was found that the average level of general vibration generated (Z-axis – Lz) depending on the frequency (F) was as follows:

table

Thus, maximum energy of vibration acceleration was observed in octave bands of 16 and 31.5 Hz, which was typical of general or household appliances vibration. For instance, such range corresponds to vibration induced by subway trains, and the 20–50 Hz frequency vibration is common to numerous industrial sources of general vibration.

Vibration acceleration was measured for 6 days, 4–6 measurements with test duration of 5 min. It was found, that during the entire observation period the measured equivalent vibration acceleration levels fluctuated in the following way:   for Х-axis within the limit of 0.7 dB; for У-axis within the limit of 2.3 dB, and for Z-axis within the limit of 1.0 dB. During one measurement session, the measured vibration acceleration fluctuations did not exceed 0.2 dB for Х-axis, 0.6 for У-axis, and 0.5 dB for Z-axis, which indicated the short-term stability of measurements appropriate for ILC.

Thus, TB designed for ILCs with measurement of general vibration factor ensures adequate simulation of the real world conditions of the general vibration measurement and makes it possible to perform measurements with sequential ILC participants’ access to TB. The use of discussed TB would enable correct comparison of the general vibration measurement results obtained by the ILC participants, and the reference laboratory. TB has been already introduced into the work of the ILC coordinator acting under the authority of the PHYSFACTOR-TEST voluntary certification system.

In case of small discontinuity of measured vibration acceleration values the expanded uncertainty can be calculated by classifying general vibration measurements as single measurements [15].

DISCUSSION

Currently, simulation of technological processes and situation at environmental objects is quite common for measurement of standardized physical factors. Therefore, the development and application of noise source physical models for the purposes of ILC are consistent with the established practice of such factors’ measurement. The need for such simulation for the purposes of ILC results from the requirements for OPT.

TBs described in this report ensure adequate simulation of acoustic noise sources at the working place and other environmental objects, and make it possible to ensure interlaboratory comparisons for all the methods applied for acoustic noise and vibration parameter measurement.

TBs contained microcontroller vibration agitator controls with feedback, which provided appropriate stability of the generated local and general vibration.

The methods developed specifically for ILCs play a crucial role in arranging and conducting ILCs. The use of standardized measurement methods makes it possible to take into account all the conditions affecting the measurement results, and provide metrologically traceable measurements during ILC [10, 11, 14]. The measurement of acoustic pressure in the ultrasonic range and general vibration acceleration is performed as direct measurement in accordance with the ILC programs, however, the methods for measuring these factors are planned to be developed and standardized.

The ILC results are evaluated based in the |En| criterion [6, 16], which ensures objectiveness of the judgment about the participating laboratory proficiency.

When performing ILC, it is possible to evaluate both the measurement results and the compliance with measurement procedure, which can be also used to assess the participants’ proficiency.

According to the comments of the participants, participation in the ILCs in accordance with the discussed technology is of great importance in terms of self-esteem and personal proficiency improvement.

CONCLUSION

Rationale for technology of acoustic noise source physical simulation, used as an object for proficiency testing during the interlaboratory comparisons, has been provided. The test benches developed meet the desired characteristics of the objects for proficiency testing, and adequately simulate the conditions at the working places and other environmental objects. When verifying the laboratory proficiency, it is necessary to assess both measurement results and measurement procedure.